Section 11: TSP command reference
Series 3700A System Switch/Multimeter Reference Manual
11-362
3700AS-901-01 Rev. D/June 2018
Also see
(on page 9-34)
(on page 11-369)
slot[slot].idn
This attribute returns a string that contains information about the plug-in card.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Not applicable
Not applicable
Not applicable
Usage
idnString
= slot[
slot
].idn
idnString
The return string
slot
Slot number (1 to 6)
Details
The information that is returned depends on whether the plug-in card in the slot is a physical card or
pseudocard.
For physical cards, this returns a comma-separated string that contains the model number,
description, firmware revision, and serial number of the plug-in card installed in the specified slot.
For pseudocards, the response is
Pseudo
, followed by the model number, description, and
???
for
the firmware revision and serial number.
Example
print(slot[3].idn)
If a 3723 is installed in slot 3, the response is:
3723,Dual 1x30 Reed Multiplexer,01.40e,1243657
Also see
None
slot[slot].interlock.override
This attribute suppresses or permits interlock errors to be generated.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Reset
Recall setup
Create configuration script
Save setup
0
Usage
value
= slot[
slot
].interlock.override
slot[
slot
].interlock.override =
value
value
Indicates the state of the interlock override:
▪
slot.ON
(
1
)
▪
slot.OFF
(
0
)
slot
The slot (1 to 6) containing the card to which the interlock state is applied