Series 3700A System Switch/Multimeter Reference Manual
Section 11: TSP command reference
3700AS-901-01 Rev. D/June 2018
11-321
scan.mode
This attribute sets the relay action when the scan starts.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Reset
Scan reset
Recall setup
Create configuration script
Save setup
0 (scan.MODE_OPEN_ALL)
Usage
scanModeSetting
= scan.mode
scan.mode =
scanModeSetting
scanModeSetting
The present scan mode setting. Set to one of the following values:
▪
scan.MODE_OPEN_ALL
or
0
▪
scan.MODE_OPEN_SELECTIVE
or
1
: See
Details
▪
scan.MODE_FIXED_ABR
or
2
: See
Details
Details
When this attribute is set to open all, channels on all slots are opened before a scan starts.
When this attribute is set to
scan.MODE_OPEN_SELECTIVE
, an intelligent open is performed.
Assuming all steps being scanned have a function value of "
nofunction
" with their DMM
configuration then:
•
All channels and analog backplane relays used in scanning are opened
•
Closed channels and backplane relays not used in scanning remain closed during the scan
If any step has a DMM configuration with a function set to any other value than
"nofunction"
:
•
Analog backplane relays 1 and 2 are opened on all slots
•
Any commonside ohms backplane relays are opened on all slots
•
Any amp channels are opened on all slots
•
All channels and backplane relays involved in scanning are opened
•
If a closed channel or backplane relay is not involved in scanning, it remains closed during the
scan
•
All channels are opened on any bank that contains backplane relays that are involved in scanning
When this attribute is set to
scan.MODE_FIXED_ABR
, it is equivalent to setting
MODE_OPEN_SELECTIVE
, except:
•
All required backplane relays are closed before the start of the scan
•
These backplane relays are not opened or closed during the scan
•
These backplane relays do not open at the end of scan
Example
scan.mode = scan.MODE_OPEN_SELECTIVE
Sets the scan mode setting to open channels
that are used in the scan.
Also see
(on page 11-324)
(on page 5-23)