G-26
Model 2790 SourceMeter® Switch System Reference Manual
Calibration program
The following calibration program is written in BASIC to help you calibrate the
Model 2790. Refer to
for more details on calibration procedures, equipment,
and connections.
Computer hardware requirements
The following computer hardware is required to run the calibration program:
•
IBM PC, AT, or compatible computer.
•
Keithley KPC-488.2, KPS-488.2, KPC-488.2AT, or CEC PC-488 IEEE-488 inter-
face for the computer.
•
Two shielded IEEE-488 connecting cables (Keithley Model 7007).
Software requirements
In order to use the calibration program, you will need the following software:
•
Microsoft QBasic (supplied with MS-DOS 5.0 or later).
•
MS-DOS version 5.0 or later.
•
HP-style Universal Language Driver, CECHP.EXE (supplied with Keithley and
CEC interface cards listed above).
Calibration equipment
The following calibration equipment is required:
•
Fluke 5700A Calibrator
•
Keithley Model 8610 Calibration Short
•
Double banana plug to BNC cables
See Section 17 for detailed equipment information and refer to these figures for
connections:
•
Low-thermal short connections:
•
DC volts and ohms connections:
•
•
Summary of Contents for 2790
Page 3: ......
Page 7: ......
Page 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Page 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Page 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Page 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Page 328: ...12 CommonCommands...
Page 336: ...13 SCPISignalOriented MeasurementCommands...
Page 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 356: ...15 SCPIReferenceTables...
Page 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Page 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Page 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Page 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Page 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 472: ...B StatusandErrorMessages...
Page 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Page 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Page 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Page 510: ...F IEEE 488BusOverview...
Page 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Page 526: ...G CalibrationReference...