17-6
Model 2790 SourceMeter
®
Switch System Reference Manual
Aborting calibration
You can abort the front panel calibration process at any time by pressing EXIT. The
instrument will then ask you to confirm your decision to abort with the following message:
ABORT CAL?
Press EXIT to abort calibration at this point or press any other key to return to the calibra-
tion process.
klqb
The Model 2790 will not respond to any remote programming commands while
the ABORT CAL? message is displayed.
Front panel calibration
Follow the steps in the following paragraphs in the order shown for comprehensive, DC-only,
and AC-only calibration procedures.
The procedures for front panel calibration include:
•
Preparing the Model 2790 for calibration
•
Front panel short and open calibration
•
DC voltage calibration
•
Resistance calibration
•
DC current calibration
•
AC voltage calibration
•
AC current calibration
•
Setting calibration dates
Preparing the Model 2790 for calibration
1.
Turn on the Model 2790 and allow it to warm up for at least two hours before per-
forming a calibration procedure.
2.
Start the calibration process as follows:
a.
Access the calibration menu by pressing SHIFT then TEST. Then display
TEST: CALIB using the up or down range key. Press ENTER.
b.
Use the up or down range key to scroll through the available calibration menu
items until the unit displays RUN, then press ENTER.
c.
At the prompt, enter the calibration code. (The default code is 002790.) Use
the left and right arrow keys to move among the digits; use the up range key to
increment numbers and press the down range key to specify alphabetic letters.
Confirm the code by pressing ENTER.
d.
Choose N at the prompt to proceed without changing the code, then press
ENTER.
Summary of Contents for 2790
Page 3: ......
Page 7: ......
Page 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Page 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Page 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Page 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Page 328: ...12 CommonCommands...
Page 336: ...13 SCPISignalOriented MeasurementCommands...
Page 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 356: ...15 SCPIReferenceTables...
Page 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Page 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Page 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Page 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Page 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 472: ...B StatusandErrorMessages...
Page 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Page 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Page 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Page 510: ...F IEEE 488BusOverview...
Page 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Page 526: ...G CalibrationReference...