8-6
Model 2790 SourceMeter
®
Switch System Reference Manual
Reading hold (autosettle)
With hold enabled (HOLD annunciator on), the first processed reading becomes the
“seed” reading and operation loops back within the device action block. After the next
reading is processed, it is checked to see if it is within the selected hold window (0.01%,
0.1%, 1%, or 10%) of the “seed” reading. If the reading is within the window, operation
again loops back within the device action block. This looping continues until the specified
number (2 to 100) of consecutive readings are within the window. If one of the readings is
not within the window, the instrument acquires a new “seed” reading and the hold process
continues.
When a hold reading is acquired, an audible beep is sounded (if enabled) and the reading
is considered a “true measurement.” The reading is held on the display until an “out of
window” reading occurs to restart the hold process.
For remote operation, the hold process seeks a new “seed” once it has been satisfied and
the reading has been released. For basic front panel operation, the hold process does not
seek a new “seed” until the held condition is removed.
klqb
Hold cannot be used when scanning (7702).
Hold example
1.
Press SHIFT and then HOLD to display the present window (0.01%, 0.1%, 1%, or
10%).
2.
To change the window, press the
or
key to display the desired window.
3.
Press ENTER. The present hold count is displayed (2 to 100).
4.
To change the hold count, use the
,
,
, and
keys to display the desired
count.
5.
Press DCV to measure DC voltage.
6.
Apply the test signal to the input of the Model 2790. Once the signal becomes stable
enough to satisfy the hold condition, the reading is released and the beeper sounds (if
enabled).
7.
Remove the hold condition by disconnecting the signal from the input. Hold will
then seek a new “seed.”
8.
To disable HOLD, press SHIFT and then HOLD.
Summary of Contents for 2790
Page 3: ......
Page 7: ......
Page 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Page 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Page 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Page 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Page 328: ...12 CommonCommands...
Page 336: ...13 SCPISignalOriented MeasurementCommands...
Page 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 356: ...15 SCPIReferenceTables...
Page 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Page 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Page 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Page 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Page 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 472: ...B StatusandErrorMessages...
Page 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Page 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Page 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Page 510: ...F IEEE 488BusOverview...
Page 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Page 526: ...G CalibrationReference...