Model 2790 SourceMeter
®
Switch System Reference Manual
6-11
f.
TRACe:POINts
2 to 55000
Set buffer size
With buffer auto-clear enabled, you can set the buffer to store from 2 to 55,000 readings. A
buffer size of zero or one is not valid (error -222).
With buffer auto-clear disabled, you cannot use this command to set buffer size (error -221)
because it is fixed at 55,000.
klqb
The full and
i
buffer full measurement events are not intended to be used with
buffer size smaller than four readings.
g.
TRACe:TSTamp:FORMat
ABSolute | DELta
Select timestamp format
For front panel operation, both timestamp formats (absolute and delta) can be recalled. For
remote programming, you can only use one timestamp at a time.
klqb
Changing the timestamp format clears the buffer.
The timestamp will only be included with a returned buffer reading if it is specified as a data
element (see FORMat:ELEMents).
h.
TRACe:FEED
SENSe | CALCulate | NONE
Select source of readings
TRACe:FEED:CONTrol
NEXT| ALWays | NEVer
Select buffer control
TRACe:FEED — The SENSe parameter selects readings before any enabled mX+b, Per-
cent, or Reciprocal math calculation. For the CALCulate parameter, the result of the calcula-
tion is stored in the buffer. The NONE parameter disables storage into the buffer. Math
functions are covered in
klqb
In order to store readings in the buffer, TRACe:FEED cannot be set to NONE.
TRACe:FEED:CONTrol — Selecting NEXT enables the buffer. After the specified number
of readings (buffer size) are stored, buffer operation disables. The ALWays parameter places
the buffer into a continuous filling mode. After the specified number of readings are stored,
operation wraps back to the first buffer location and overwrites the previous readings. The
NEVer parameter disables buffer operation.
i.
TRACe:DATA?
Read buffer
Use TRACE:DATA? to retrieve all readings that are stored in the buffer. You can send this
command even if the instrument is still storing readings. When TRACe:DATA? is sent, it
will return the readings stored up to that point in time. Subsequent TRACe:DATA? com-
mands will not retrieve readings already returned. However, once the buffer has filled and
you have retrieved all buffer readings, you can again send TRACe:DATA? to retrieve all the
stored readings.
The data elements returned with each stored reading depends on which ones are selected
with FORMat:ELEMents command (see
Summary of Contents for 2790
Page 3: ......
Page 7: ......
Page 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Page 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Page 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Page 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Page 328: ...12 CommonCommands...
Page 336: ...13 SCPISignalOriented MeasurementCommands...
Page 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Page 356: ...15 SCPIReferenceTables...
Page 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Page 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Page 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Page 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Page 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Page 472: ...B StatusandErrorMessages...
Page 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Page 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Page 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Page 510: ...F IEEE 488BusOverview...
Page 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Page 526: ...G CalibrationReference...