9-18
Limits and Digital I/O
Model 2750 Multimeter/Switch System User’s Manual
Remote Operation
— For remote operation, make sure both Limit 1 and Limit 2 are
enabled. The following table evaluates the three possible pass/fail combinations for this
example.
Keep in mind that a fail condition must be reset before testing the next resistor. Fail can be
reset manually or automatically (see
, CLEar command).
Digital outputs
With the digital outputs of the Model 2750 enabled, the digital outputs will respond as fol-
lows for each resistor reading:
Limit 1 result
Limit 2 result
Resistor tolerance
Bin assignment
Pass
Pass
>
1%
1
Fail Pass
>
5%
2
Fail Fail >5%
3
LO limit 2
LO limit 1
HI limit 1
HI limit 2
Resistor
tolerance
Bin
Affected
outputs*
Pass
Pass
Pass
Pass
1%
1
None
Pass
Fail
Pass
Pass
5%
2
#1 and #5
Pass
Pass
Fail
Pass
5%
2
#2 and #5
Fail
Fail
Pass
Pass
>5%
3
#1, #3 and #5
Pass
Pass
Fail
Fail
>5%
3
#2, #4 and #5
*Affected outputs are pulled (or pulsed) high or low when a limit test fails.
2750-900-01.book Page 18 Wednesday, August 3, 2011 7:56 AM
Summary of Contents for 2750
Page 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Page 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Page 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Page 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...