Model 2750 Multimeter/Switch System User’s Manual
Basic DMM Operation
3-41
Figure 3-13
Continuity connections
Continuity testing procedure
NOTE
Make sure the INPUTS switch is in the correct position. To use front panel
inputs, it must be in the “F” (out) position. For switching modules, it must be in
the “R” (in) position.
1.
Apply the resistance to be tested, and, if using a switching module, close the appro-
priate channel.
2.
Press SHIFT and then CONT to display the present threshold LEVEL.
3.
Use
,
,
Δ
, and
∇
to key in the desired level (1 to 1000
Ω
), and press ENTER.
4.
If the measurement is below the threshold level, the beeper will sound and the
ohms reading will be displayed. If the measurement is above the threshold level,
the message “OPEN” will be displayed.
5.
To disable continuity testing, select a different function (i.e., press DCV).
NOTE
The beeper can be disabled using the SYSTem:BEEPer:STATe OFF command.
However, the beeper will automatically enable the next time the continuity test-
ing function is selected.
Limits and digital outputs cannot be used when testing continuity with the conti-
nuity (CONT) function. If you need to use these operations, use the
Ω
2 function
to test continuity.
H
L
CH 1-20
Model 7700
Switching Module
Resistance
Under Test
Note: Source current flows from input high to input low.
Input HI
Input LO
Resistance
Under Test
A. Front Panel Connections
B. Model 7700 Connections
Model 2750
2750-900-01.book Page 41 Wednesday, August 3, 2011 7:56 AM
Summary of Contents for 2750
Page 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Page 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Page 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Page 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...