Model 2750 Multimeter/Switch System User’s Manual
Basic DMM Operation
3-25
Temperature measurements
The Model 2750 can measure temperature using thermocouples, thermistors, and 4-wire
RTDs. When deciding which temperature sensor to use, keep in mind that the thermocou-
ple is the most versatile, the thermistor is the most sensitive, and the 4-wire RTD is the
most stable.
Thermocouples
For thermocouples, temperature measurement range depends on which type of thermocou-
ple is being used. Thermocouples that are supported include types J, K, N, T, E, R, S, and
B.
When two wires made up of dissimilar metals are joined together, a voltage is generated.
The generated voltage is a function of temperature. As temperature changes, the voltage
changes. The thermocouple voltage equates to a temperature reading. This is the basic
operation principle of the thermocouple.
NOTE
The equation to calculate thermocouple temperature is provided in
When you connect a thermocouple directly to the input of the Model 2750, at least one of
those connections will be a junction made up of two dissimilar metals. Hence, another
voltage is introduced and is algebraically added to the thermocouple voltage. The result
will be an erroneous temperature measurement.
To cancel the affects of the unwanted thermal voltage, the thermocouple circuit requires a
reference junction that is at a known temperature.
Type
Range
Resolution
J
-200°C to 760°C
0.001°C
K
-200°C to 1372°C 0.001°C
N
-200°C to 1300°C 0.001°C
T
-200°C to 400°C
0.001°C
E
-200°C to 1000°C 0.001°C
R
0°C to 1768°C
0.1°C
S
0°C to 1786°C
0.1°C
B
+350°C to 1820°C 0.1°C
2750-900-01.book Page 25 Wednesday, August 3, 2011 7:56 AM
Summary of Contents for 2750
Page 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Page 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Page 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Page 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...