![Keithley 2750 Service Manual Download Page 168](http://html1.mh-extra.com/html/keithley/2750/2750_service-manual_3957779168.webp)
C-2
Calibration Program
Model 2750 Multimeter/Switch System Service Manual
Introduction
This appendix includes a calibration program written in BASIC to help you calibrate the
Model 2750. Refer to
for more details on calibration procedures, equipment, and
connections.
Computer hardware requirements
The following computer hardware is required to run the calibration program:
•
IBM PC, AT, or compatible computer.
•
Keithley KPC-488.2, KPS-488.2, or KPC-488.2AT, or CEC PC-488 IEEE-488
interface for the computer.
•
Two shielded IEEE-488 connecting cables (Keithley Model 7007).
Software requirements
In order to use the calibration program, you will need the following software:
•
Microsoft QBasic (supplied with MS-DOS 5.0 or later).
•
MS-DOS version 5.0 or later.
•
HP-style Universal Language Driver, CECHP.EXE (supplied with Keithley and
CEC interface cards listed above).
Calibration equipment
The following calibration equipment is required:
•
Fluke 5700A Calibrator.
•
Keithley Model 8610 Calibration Short.
•
Double banana plug to BNC cables.
for detailed equipment information, and refer to these figures for
connections:
•
Low-thermal short connections:
.
•
DC volts and ohms connections:
•
DC amps and AC amps connections:
•
.
Summary of Contents for 2750
Page 13: ...1 PerformanceVerification...
Page 48: ...1 36 Performance Verification Model 2750 Multimeter Switch System Service Manual...
Page 49: ...2 Calibration...
Page 74: ...2 26 Calibration Model 2750 Multimeter Switch System Service Manual...
Page 75: ...Routine Maintenance 3 RoutineMaintenance...
Page 86: ...3 12 Routine Maintenance Model 2750 Multimeter Switch System Service Manual...
Page 87: ...Troubleshooting 4 Troubleshooting...
Page 111: ...Disassembly 5 Disassembly...
Page 118: ...5 8 Disassembly Model 2750 Multimeter Switch System Service Manual...
Page 119: ...Replaceable Parts 6 ReplaceableParts...
Page 135: ...Specifications A AccuracyandOptimization...
Page 140: ...A 6 Accuracy and Optimization Model 2750 Multimeter Switch System Service Manual...
Page 141: ...Calibration Reference B CalibrationReference...
Page 167: ...Calibration Program C CalibrationProgram...
Page 178: ......
Page 179: ......