Model 2700 Multimeter/Switch System User’s Manual
Basic DMM Operation
3-7
Throughout this manual, you will encounter commands that can use the <clist> parameter.
The <clist> simply indicates that the associated command can be used to configure a scan
channel. For example:
SENSe:FUNCtion 'VOLTage:AC'
' Select ACV function.
SENSe:FUNCtion 'VOLTage:AC',(@101)
' Configure scan channel 101 for ACV.
While in the normal measurement display state, the first command simply selects the ACV
function. The second command configures channel 101 to measure ACV when it is
scanned.
for detailed information on scanning.
Voltage measurements (DCV and ACV)
The Model 2700 can make DCV measurements from 0.1µV to 1000V and ACV
measurements from 0.1µV to 750V RMS, 1000V peak.
DCV input resistance:
100V and 1000V ranges: 10M
Ω
100mV, 1V, and 10V ranges: >10G
Ω
|| <400pF or 10M
Ω
ACV input impedance:
1M
Ω
|| <100pF
DCV input divider
Normally, the input resistance for the 100mVDC, 1VDC, and 10VDC ranges is >10G
Ω
,
while the input resistance of the 100VDC and 1000VDC ranges is 10M
Ω
. However, the
input resistance for the three lower DCV ranges can also be set to 10M
Ω
by enabling the
input divider.
With the input resistance lowered, a more stable 0V reading is achieved with an open
input. Also, some external devices (such as a high voltage probe) must be terminated to a
10M
Ω
load.
The input divider cannot be enabled from the front panel. For remote programming, the
following command controls the input divider:
VOLT:IDIVider <b> ' Enable (ON) or disable (OFF) the DCV input divider.
Summary of Contents for 2700
Page 3: ......
Page 7: ......
Page 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Page 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Page 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Page 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Page 336: ...12 CommonCommands...
Page 344: ...13 SCPISignalOriented MeasurementCommands...
Page 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Page 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Page 364: ...15 SCPIReferenceTables...
Page 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Page 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Page 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 400: ...B Model7700ConnectionGuide...
Page 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Page 412: ...C StatusandErrorMessages...
Page 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Page 420: ...D SignalProcessing SequenceandDataFlow...
Page 434: ...E MeasurementConsiderations...
Page 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Page 454: ...G IEEE 488BusOverview...
Page 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Page 470: ...H KE2700InstrumentDriver Examples...
Page 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Page 499: ......