Model 2700 Multimeter/Switch System User’s Manual
SCPI Reference Tables
15-3
Table 15-1
CALCulate command summary
Command
Description
Default
parameter
Ref
SCPI
CALCulate[1]
Subsystem to control CALC 1:
✓
:FORMat <name>
[<, clist>]
Select math format (NONE, MXB, PERCent, or
RECiprocal).
PERCent
✓
:FORMat? [<clist>]
Query math format.
✓
:KMATh
Path to configure math calculations:
:MMFactor <NRf>
[, <clist>]
Set “m” factor for mx+b (-4294967295 to
+4294967295).
1
1
:MA1Factor <NRf>
[, <clist>]
Set “m” factor for mx+b (-4294967295 to
+4294967295).
1
1
:MMFactor?
[<clist>]
Query “m” factor.
1
:MA1Factor?
[<clist>]
Query “m” factor.
1
:MBFactor <NRf>
[, <clist>]
Set “b” factor for mx+b (-4294967295 to
+4294967295).
2
0
:MA0Factor <NRf>
[, <clist>]
Set “b” factor for mx+b (-4294967295 to
+4294967295).
2
0
:MBFactor? [<clist>]
Query “b” factor.
2
:MA0Factor?
[<clist>]
Query “b” factor.
2
:MUNits <char>
Specify units for mx+b reading.
3
‘X’
:MUNits?
Query “mx+b” units.
:PERCent <NRf>
[, <clist>]
Set target value for PERCent calculation
(-4294967295 to +4294967295).
1
:ACQuire
Use input signal as target value.
:PERCent? [<clist>]
Query percent.
:STATe <b> [, <clist>]
Enable or disable kmath calculation.
0
✓
:STATe? [< clist>]
Query state of kmath function.
✓
:DATA?
Read result of kmath calculation.
✓
CALCulate2
Subsystem to control CALC 2:
✓
:FORMat <name>
Select math format: (MEAN, SDEViation,
MAXimum, MINimum, PKPK, or NONE).
NONE
✓
:FORMat?
Query math format.
✓
:STATe <b>
Enable or disable calculation.
0
✓
:STATe?
Query state of math function.
✓
:IMMediate
Recalculate raw input data in buffer.
✓
:IMMediate?
Perform calculation and read result.
✓
:DATA?
Read math result of CALC 2.
✓
Summary of Contents for 2700
Page 3: ......
Page 7: ......
Page 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Page 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Page 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Page 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Page 336: ...12 CommonCommands...
Page 344: ...13 SCPISignalOriented MeasurementCommands...
Page 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Page 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Page 364: ...15 SCPIReferenceTables...
Page 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Page 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Page 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 400: ...B Model7700ConnectionGuide...
Page 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Page 412: ...C StatusandErrorMessages...
Page 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Page 420: ...D SignalProcessing SequenceandDataFlow...
Page 434: ...E MeasurementConsiderations...
Page 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Page 454: ...G IEEE 488BusOverview...
Page 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Page 470: ...H KE2700InstrumentDriver Examples...
Page 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Page 499: ......