14-6
FORMat and Misc SYSTem Commands
Model 2700 Multimeter/Switch System User’s Manual
FORMat:ELEMents <item list>
Parameters
READing
= DMM reading
UNITs
= Units
TSTamp
= Timestamp
RNUMber
= Reading number
CHANnel
= Channel number
LIMits
= Limits reading
The specified elements are included in the data string in response to FETCh?, READ?,
MEASure?, and TRACe:DATA?. Note that each element in the item list must be separated
by a comma (i.e. send “FORMat:ELEMents READing, UNITs, TSTamp, RNUMber,
CHANnel, LIMits” to include all elements in the data string). The elements for the ASCii
format are shown in
.
An overflow reading is returned as +9.9E37. When a specified data element has invalid
data associated with it, NAN (Not A Number) will be the response. NAN is returned as
+9.9E37.
Timestamp
— Timestamp references the returned data string to a point in time. There are
two timestamps: relative and real-time clock. The following command selects the
timestamp:
SYSTem:TSTamp:TYPE <name>
' Select timestamp type; RELative or RTCL.
NOTE
The real-time clock timestamp can only be returned for the ASCII data format.
For the binary formats, timestamp will not be sent with the real-time clock
selected.
The relative timestamp operates as a timer that starts at zero seconds when the instrument
is turned on, or when the relative timestamp is reset (SYSTem:TSTamp:RELative:RESet).
After 99,999.99 seconds, the timer resets back to zero and starts over.
For buffer readings recalled from the front panel, the relative timestamp is referenced to
the first reading stored in the buffer (absolute format) which is timestamped at 0 seconds,
and to the time between each stored reading (delta format). For remote programming, you
can only return the absolute or delta timestamp. The following command is used to select
relative timestamp format for the buffer:
TRACe:TSTamp:FORMat
' Select timestamp format; ABSolute or DELTa.
Reading number
— The reading counter starts at zero when the Model 2700 is turned on.
When returning buffer readings using TRACe:DATA?, each reading will be referenced to
the first reading, which is #0. The following command will reset the counter:
SYSTem:RNUMber:RESet.
Channel number
— The channel number indicates the switching module channel for the
reading. Channel number 000 indicates that no channel was closed.
Summary of Contents for 2700
Page 3: ......
Page 7: ......
Page 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Page 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Page 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Page 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Page 336: ...12 CommonCommands...
Page 344: ...13 SCPISignalOriented MeasurementCommands...
Page 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Page 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Page 364: ...15 SCPIReferenceTables...
Page 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Page 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Page 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 400: ...B Model7700ConnectionGuide...
Page 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Page 412: ...C StatusandErrorMessages...
Page 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Page 420: ...D SignalProcessing SequenceandDataFlow...
Page 434: ...E MeasurementConsiderations...
Page 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Page 454: ...G IEEE 488BusOverview...
Page 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Page 470: ...H KE2700InstrumentDriver Examples...
Page 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Page 499: ......