10-20
Remote Operations
Model 2700 Multimeter/Switch System User’s Manual
Selecting and configuring RS-232 interface
After selecting (enabling) the RS-232 interface, you will then set the baud rate, flow
control, and terminator.
1.
Press the SHIFT key and then the RS-232 key. The RS 232 ON or RS 232 OFF
message will be displayed.
2.
If the RS-232 is already ON, press ENTER and proceed to step 3. Otherwise, press
the
key to place the cursor on OFF, press the
Δ
or
∇
key to display the ON
state, and then press ENTER.
NOTE
Enabling (ON) the RS-232 interface disables (OFF) the GPIB. Disabling the
RS-232 interface enables the GPIB.
3.
To retain the presently displayed BAUD rate, press ENTER and proceed to step 4.
Otherwise, press the
key to place the cursor on the baud rate value, use the
Δ
or
∇
key to display the desired baud rate, and then press ENTER.
4.
To retain the presently displayed FLOW control, press ENTER and proceed to step
5. Otherwise, press the
key to place the cursor on the flow control setting, use
the
Δ
or
∇
key to display the alternate flow control setting, and then press
ENTER.
5.
To retain the presently displayed terminator (Tx TERM), press ENTER. Otherwise,
press the
key to place the cursor on the terminator setting, use the
Δ
or
∇
key to
display the desired terminator, and then press ENTER.
RS-232 connections
The RS-232 serial port is connected to the serial port of a computer using a straight-
through RS-232 cable terminated with DB-9 connectors.
Do not use a null modem cable
.
The serial port uses the transmit (TXD), receive (RXD), and signal ground (GND) lines of
the RS-232 standard. It does not use the hardware handshaking lines, CTS and RTS.
shows the rear panel connector for the RS-232 interface, and
shows the pinout for the connector.
If your computer uses a DB-25 connector for the RS-232 interface, you will need a cable
or adapter with a DB-25 connector on one end and a DB-9 connector on the other, wired
straight through (not null modem).
Summary of Contents for 2700
Page 3: ......
Page 7: ......
Page 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Page 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Page 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Page 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Page 336: ...12 CommonCommands...
Page 344: ...13 SCPISignalOriented MeasurementCommands...
Page 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Page 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Page 364: ...15 SCPIReferenceTables...
Page 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Page 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Page 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 400: ...B Model7700ConnectionGuide...
Page 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Page 412: ...C StatusandErrorMessages...
Page 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Page 420: ...D SignalProcessing SequenceandDataFlow...
Page 434: ...E MeasurementConsiderations...
Page 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Page 454: ...G IEEE 488BusOverview...
Page 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Page 470: ...H KE2700InstrumentDriver Examples...
Page 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Page 499: ......