Model 2700 Multimeter/Switch System User’s Manual
Range, Digits, Rate, Bandwidth, and Filter
4-11
Table 4-5
Rate and bandwidth commands
Commands
1, 7
Description
Default
Integration rate commands
[SENSe[1]]
Optional root command.
:VOLTage[:DC]:NPLCycles <n> [, <clist>]
Set rate for DCV in PLCs; <n> = 0.01 to x
2
.
5.0
:VOLTage[:DC]:APERture <n> [, <clist>]
Set rate for DCV in secs; <n> = x to 1
3
.
(Note 4)
:VOLTage:AC:NPLCycles <n> [, <clist>]
Set rate for ACV in PLCs; <n> = 0.01 to xx
2,5
. 5.0
:VOLTage:AC:APERture <n> [, <clist>]
Set rate for ACV in secs; <n> = x to 1
3,5
.
(Note 4)
:CURRent[:DC]:NPLCycles <n> [, <clist>]
Set rate for DCI in PLCs; <n> = 0.01 to x
2
.
5.0
:CURRent[:DC]:APERture <n> [, <clist>]
Set rate for DCI in secs; <n> = x to 1
3
.
(Note 4)
:CURRent:AC:NPLCycles <n> [, <clist>]
Set rate for ACI in PLCs; <n> = 0.01 to x
2,5
.
5.0
:CURRent:AC:APERture <n> [, <clist>]
Set rate for ACI in secs; <n> = x to 1
3,5
.
(Note 4)
:RESistance:NPLCycles <n> [, <clist>]
Set rate for
Ω
2 in PLCs; <n> = 0.01 to x
2
.
5.0
:RESistance:APERture <n> [, <clist>]
Set rate for
Ω
2 in secs; <n> = x to 1
3
.
(Note 4)
:FRESistance:NPLCycles <n> [, <clist>]
Set rate for
Ω
4 in PLCs; <n> = 0.01 to x
2
.
5.0
:FRESistance:APERture <n> [, <clist>]
Set rate for
Ω
4 in secs; <n> = x to 1
3
.
(Note 4)
:TEMPerature:NPLCycles <n> [, <clist>]
Set rate for TEMP in PLCs; <n> = 0.01 to x
2
.
5.0
:TEMPerature:APERture <n> [, <clist>]
Set rate for TEMP in secs; <n> = x to 1
3
.
(Note 4)
Bandwidth commands
[SENSe[1]]
Optional root command.
:VOLTage:AC:DETector:BANDwidth
<NRf> [, <clist>]
Set AC bandwidth for ACV in Hertz; <NRf> =
3 to 3e5
6
.
30
:CURRent:AC:DETector:BANDwidth
<NRf> [, <clist>]
Set AC bandwidth for ACI in Hertz;
<NRf> = 3 to 3e5
6
.
30
Channel list parameter:
<clist> = (@SCH)
where: S =
Mainframe slot number (1 or 2)
CH = Switching module channel number (must be 2 digits)
Examples:
(@101) = Slot 1, Channel 1
(@101, 203) = Slot 1, Channel 1 and Slot 2, Channel 3
(@101:110) = Slot 1, Channels 1 through 10
Notes:
1. The <clist> parameter is used to configure one or more channels for a scan. Each channel in the <clist> must be set to the function
specified by the command. If not, a conflict error (-221) will occur. For example, RESistance:NPLCycles 1, (@101) is only valid if
scan channel 101 is set for the
Ω
2 function.
2. For 60Hz line power, x = 60. For 50Hz line power, x = 50.
3. For 60Hz line power, x = 1.67e-4. For 50Hz line power, x = 2e-4.
4. For 60Hz line power, the default is 16.67msec. For 50Hz line power, the default is 20msec.
5. Commands to set rate for ACV and ACI are only valid if bandwidth is set to 300 (300Hz to 300kHz). See
, for details.
6. The instrument will actually accept a parameter value up to 10e6, but it will default to 3e5.
7. [:DC] is optional for the commands to set DCV and DCI integration rate.
Summary of Contents for 2700
Page 3: ......
Page 7: ......
Page 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Page 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Page 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Page 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Page 336: ...12 CommonCommands...
Page 344: ...13 SCPISignalOriented MeasurementCommands...
Page 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Page 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Page 364: ...15 SCPIReferenceTables...
Page 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Page 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Page 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 400: ...B Model7700ConnectionGuide...
Page 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Page 412: ...C StatusandErrorMessages...
Page 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Page 420: ...D SignalProcessing SequenceandDataFlow...
Page 434: ...E MeasurementConsiderations...
Page 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Page 454: ...G IEEE 488BusOverview...
Page 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Page 470: ...H KE2700InstrumentDriver Examples...
Page 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Page 499: ......