Model 2700 Multimeter/Switch System User’s Manual
Basic DMM Operation
3-19
6.
Observe the displayed reading. If the “OVERFLOW” message is displayed, select
a higher range until a normal reading is displayed (or press AUTO for
autoranging). For manual ranging, use the lowest possible range for the best
resolution.
7.
To measure another amps channel, repeat steps 5 and 6.
8.
When finished, press OPEN if there is a channel closed.
NOTE
When you have an amps-only channel closed, you cannot select a non-amps
function. For example, if channel 21 of the Model 7700 is closed, you cannot
select the DCV function (“INVALID FUNC” displayed).
AMPS fuse replacement (front panel AMPS input)
WARNING
Make sure the instrument is disconnected from the power line and
other equipment before replacing the AMPS fuse.
1.
Turn off the power and disconnect the power line and test leads.
2.
From the front panel, gently push in the AMPS fuse holder with a flat blade
screwdriver and rotate the fuse holder one-quarter turn counterclockwise.
3.
Remove the fuse and replace it with the same type (3A, 250V, fast-blow, 5
×
20mm). The Keithley part number is FU-99-1.
CAUTION
Do not use a fuse with a higher current rating than specified or
instrument damage may occur. If the instrument repeatedly blows
fuses, locate and correct the cause of the trouble before replacing the
fuse.
4.
Install the new fuse by reversing the procedure above.
NOTE
For the Model 7700 switching module and other similar modules that support
the amps function, there are solder mount amps fuses. See the Model 2700
Service Manual for fuse replacement information.
Summary of Contents for 2700
Page 3: ......
Page 7: ......
Page 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Page 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Page 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Page 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Page 336: ...12 CommonCommands...
Page 344: ...13 SCPISignalOriented MeasurementCommands...
Page 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Page 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Page 364: ...15 SCPIReferenceTables...
Page 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Page 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Page 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Page 400: ...B Model7700ConnectionGuide...
Page 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Page 412: ...C StatusandErrorMessages...
Page 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Page 420: ...D SignalProcessing SequenceandDataFlow...
Page 434: ...E MeasurementConsiderations...
Page 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Page 454: ...G IEEE 488BusOverview...
Page 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Page 470: ...H KE2700InstrumentDriver Examples...
Page 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Page 499: ......