Model 2657A High Power System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2657A-901-01 Rev. B/December 2012
7-51
Also see
(on page 7-54)
digio.trigger[N].clear()
This function clears the trigger event on a digital I/O line.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function Yes
Usage
digio.trigger[
N
].clear()
N
Digital I/O trigger line (1 to 14)
Details
The event detector of a trigger enters the detected state when an event is detected. It is cleared when
digio.trigger[
N
].wait()
or
digio.trigger[
N
].clear()
is called.
digio.trigger[
N
].clear()
clears the event detector of the specified trigger line, discards the history of the
trigger line, and clears the
digio.trigger[
N
].overrun
attribute.
Example
digio.trigger[2].clear()
Clears the trigger event detector on I/O line 2.
Also see
(on page 7-53)
(on page 7-57)
digio.trigger[N].EVENT_ID
This constant identifies the trigger event generated by the digital I/O line N.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Constant Yes
Usage
eventID
= digio.trigger[
N
].EVENT_ID
eventID
The trigger event number
N
Digital I/O trigger line (1 to 14)
Details
To have another trigger object respond to trigger events generated by the trigger line, set the other object's
stimulus attribute to the value of this constant.
Example 1
digio.trigger[5].stimulus =
digio.trigger[3].EVENT_ID
Uses a trigger event on digital I/O
trigger line 3 to be the stimulus for
digital I/O trigger line 5.