Appendix B: Calibration
Model 2651A High Power System SourceMeter® Instrument Reference Manual
B-16
2651A-901-01 Rev. A / March 2011
Adjustment
Introduction
Use the procedures in this section to calibrate the Keithley Instruments Model 2651A High Power
System SourceMeter
®
Instrument. These procedures require accurate test equipment to measure
precise DC voltages and currents.
CAUTION
The information in this section is intended for qualified service personnel only. Do not attempt these
procedures unless you are qualified to do so.
NOTE
Product specifications are subject to change. Listed uncertainties and test limits are provided only as
an example. Always verify values against actual product specifications.
Environmental conditions
Temperature and relative humidity
Conduct the calibration procedures at an ambient temperature of 18° C to 28° C (65° F to 82° F), with
relative humidity of less than 70 percent (unless otherwise noted).
NOTE
Product specifications that are listed as 18º C to 28º C assume adjustment has been done at 23º C. If
the Model 2651A is adjusted at a different temperature, the specifications apply to +/-5º C of that
adjustment temperature.
Line power
The Model 2651A requires a line voltage of 100 V to 240 V at line frequency of 50 Hz or 60 Hz. The
instrument must be calibrated within this range.
Warm-up period
Allow the Model 2651A to warm up for at least two hours before performing calibration.
If the instrument has been subjected to temperature extremes (those outside the ranges stated
above), allow additional time for the instrument's internal temperature to stabilize. Typically, allow one
extra hour to stabilize a unit that is 10°C (18°F) outside the specified temperature range.
Also, allow the test equipment to warm up for the minimum time specified by the manufacturer.