Section 7: Command reference
Model 2651A High Power System SourceMeter® Instrument Reference Manual
7-338
2651A-901-01 Rev. A / March 2011
Details
This function waits for an event blender trigger event. If one or more trigger events were detected since the last
time
trigger.blender[
N
].wait()
or
trigger.blender[
N
].clear()
was called, this function returns
immediately.
After detecting a trigger with this function, the event detector automatically resets and rearms. This is true
regardless of the number of events detected.
Example
digio.trigger[3].mode = digio.TRIG_FALLING
digio.trigger[5].mode = digio.TRIG_FALLING
trigger.blender[1].orenable = true
trigger.blender[1].stimulus[1] = digio.trigger[3].EVENT_ID
trigger.blender[1].stimulus[2] = digio.trigger[5].EVENT_ID
print(trigger.blender[1].wait(3))
Generate a trigger blender 1
event when a digital I/O
trigger happens either on
line 3 or 5.
Wait three seconds while
checking if trigger blender 1
event has occurred.
If the blender trigger event
has happened, then
true
is
output. If the trigger event
has not happened, then
false
is output after the
timeout expires.
Also see
(on page 7-333)
trigger.clear()
This function clears the command interface trigger event detector.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function No
Usage
trigger.clear()
Details
The trigger event detector indicates if an event has been detected since the last
trigger.wait()
call. This
function clears the trigger's event detector and discards the previous history of command interface trigger events.
Also see