![Keithley 2651A Reference Manual Download Page 522](http://html1.mh-extra.com/html/keithley/2651a/2651a_reference-manual_661729522.webp)
Section 7: Command reference
Model 2651A High Power System SourceMeter® Instrument Reference Manual
7-214
2651A-901-01 Rev. A / March 2011
Usage
sDelay
= smu
X
.source.delay
smu
X
.source.delay =
sDelay
sDelay
Set to your desired source delay value. For example, to specify an additional 10 ms
source delay, set the value to
0.010
.
Alternatively, you can set it one of the following values:
0
or
smu
X
.DELAY_OFF
: No delay
-1
or
smu
X
.DELAY_AUTO
: Automatic delay value
X
SMU channel (for example,
smua.source.delay
applies to SMU Channel A).
Details
This attribute allows for additional delay (settling time) after an output step. Set
sDelay
to a user-defined value
(in seconds). Alternatively, set
sDelay
to
smu
X
.DELAY_OFF
or
smuX.DELAY_AUTO
.
The
smu
X
.DELAY_AUTO
setting causes a range-dependent delay to be inserted when the source is changed.
Range dependent delays are based on the output settling time values in the instrument's specifications.
Example
smua.source.delay = smua.DELAY_AUTO
Sets the delay for SMU Channel A
to automatic (a range dependent
delay is inserted when ever the
source is changed).
Also see
(on page 7-194)
(on page 7-209)
smuX.source.func
This attribute sets the source function (V source or I source).
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
SMU reset
Instrument reset
Recall
Saved setups
smuX.OUTPUT_DCVOLTS
Usage
sFunction
= smu
X
.source.func
smu
X
.source.func =
sFunction
sFunction
The source function. Set to one of the following values:
0
or
smu
X
.OUTPUT_DCAMPS
: Selects current source function.
1
or
smu
X
.OUTPUT_DCVOLTS
: Selects voltage source function.
X
SMU channel (for example,
smua.source.func
applies to SMU Channel A).