Model 2651A High Power System SourceMeter® Instrument Reference Manual
Section 7: Command reference
2651A-901-01 Rev. A / March 2011
7-171
setup.save()
This function saves the present setup as a user-saved setup.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function Yes
Usage
setup.save(
id
)
id
An integer or string specifying where to save the user setup:
•
Save in nonvolatile memory (
1
to
5
)
•
Save as user-saved setup on a USB flash drive (
"/path/filename"
)
Details
When the
id
parameter is an integer (
n
), it is interpreted as the setup number to save to the instrument's
nonvolatile memory.
NOTE
When you save to a specified integer (
1
to
5
) in nonvolatile memory, the previous setup at that same
location is overwritten.
When the
id
parameter is a string, it is interpreted as the path and file name of the location to save the present
setup on a USB flash drive. The path may be absolute or relative to the current working directory.
Example
setup.save(5)
Saves the present setup at location 5.
Also see
smuX.abort()
This function terminates all overlapped operations on the specified SMU.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function Yes
Usage
smu
X
.abort()
X
Source-measure unit (SMU) channel (for example,
smua.abort()
specifies SMU
channel A).
Details
smu
X
.abort()
will not turn the output off or change any settings.
If used to abort a sweep, when executed, the SMU will exit its trigger model immediately and return to the trigger
model's idle state.