Section 2: General operation
Model 2651A High Power System SourceMeter® Instrument Reference Manual
2-8
2651A-901-01 Rev. A / March 2011
Turning your instrument on and off
Procedure
Follow the procedure below to connect the Model 2651A to line power and turn on the instrument.
The Model 2651A operates from a line voltage of 100 V to 240 V at a frequency of 50 Hz or 60 Hz.
Line voltage is automatically sensed (there are no switches to set). Make sure the operating voltage
in your area is compatible.
CAUTION
Operating the instrument on an incorrect line voltage may cause damage to the instrument, possibly
voiding the warranty.
To turn a Model 2651A on and off:
1. Before plugging in the power cord, make sure that the front panel POWER switch is in the off (O)
position.
2. Connect the female end of the supplied power cord to the AC receptacle on the rear panel.
3. Connect the other end of the power cord to a grounded AC outlet.
WARNING
The power cord supplied with the Model 2651A contains a separate ground wire for use with
grounded outlets. When proper connections are made, the instrument chassis is connected to
power-line ground through the ground wire in the power cord. Failure to use a grounded outlet
may result in personal injury or death due to electric shock.
4. To turn your instrument on, press the front panel
POWER
switch to place it in the on (I) position.
5. To turn your instrument off, press the front panel
POWER
switch to place it in the off (O) position.
Warm-up
The Model 2651A must be turned on and allowed to warm up for at least two hours to achieve rated
accuracies.
Line frequency configuration
The factory configures the Model 2651A to automatically detect the line frequency (either 50 Hz or
60 Hz) at each power-up. This detected line frequency will be used for aperture (NPLC) calculations.
In noisy environments, you can manually configure the unit to match the actual line frequency.