2606B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2606B-901-01 Rev. B / May 2018
7-345
Example
digio.trigger[3].mode = digio.TRIG_FALLING
digio.trigger[5].mode = digio.TRIG_FALLING
trigger.blender[1].orenable = true
trigger.blender[1].stimulus[1] = digio.trigger[3].EVENT_ID
trigger.blender[1].stimulus[2] = digio.trigger[5].EVENT_ID
print(trigger.blender[1].wait(3))
Generate a trigger blender 1
event when a digital I/O
trigger happens either on
line 3 or 5.
Wait three seconds while
checking if trigger blender 1
event has occurred.
If the blender trigger event
has happened, then
true
is
output. If the trigger event
has not happened, then
false
is output after the
timeout expires.
Also see
(on page 7-340)
trigger.clear()
This function clears the command interface trigger event detector.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
No
Usage
trigger.clear()
Details
The trigger event detector indicates if a trigger event has been detected since the last
trigger.wait()
call.
trigger.clear()
clears the trigger event detector and discards the history
of command interface trigger events.
Also see