Section 7: TSP command reference
2606B System SourceMeter® Instrument Reference Manual
7-262
2606B-901-01 Rev. B / May 2018
Also see
(on page E-21)
status.measurement.voltage_limit.*
This attribute contains the measurement event voltage limit summary register set.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute
- -
- -
- -
- -
.condition (R)
Yes
Not applicable
Not saved
Not applicable
.enable (RW)
Yes
Status reset
Not saved
0
.event (R)
Yes
Status reset
Not saved
0
.ntr (RW)
Yes
Status reset
Not saved
0
.ptr (RW)
Yes
Status reset
Not saved
6 (All bits set)
Usage
measurementRegister
= status.measurement.voltage_limit.condition
measurementRegister
= status.measurement.voltage_limit.enable
measurementRegister
= status.measurement.voltage_limit.event
measurementRegister
= status.measurement.voltage_limit.ntr
measurementRegister
= status.measurement.voltage_limit.ptr
status.measurement.voltage_limit.enable =
measurementRegister
status.measurement.voltage_limit.ntr =
measurementRegister
status.measurement.voltage_limit.ptr =
measurementRegister
measurementRegister
The status of the measurement voltage limit summary register; a zero (0)
indicates no bits set (also send 0 to clear all bits); other decimal values
indicate various bit settings
Details
These attributes read or write to the measurement event voltage limit summary registers. Reading a
status register returns a value. The binary equivalent of the returned value indicates which register
bits are set. The least significant bit of the binary number is bit B0, and the most significant bit is bit
B15.
For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to
Enable and transition registers
(on page E-16). The individual bits of this
register are defined in the following table.
Bit
Value
Description
B0
Not used
Not applicable.
B1
status.measurement.voltage_limit.SMUA
Set bit indicates the enabled VLMT bit for
the SMU A measurement register is set.
Bit B1 decimal value: 2
Binary value: 0000 0010
B2
status.measurement.voltage_limit.SMUB
Set bit indicates the enabled VLMT bit for
the SMU B measurement register is set.
Bit B2 decimal value: 4
Binary value: 0000 0100
B3-B15
Not used
Not applicable.