Section 7: TSP command reference
2606B System SourceMeter® Instrument Reference Manual
7-212
2606B-901-01 Rev. B / May 2018
smuX.source.autorangeY
This attribute contains the state of the source autorange control (on/off).
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
SMU reset
Instrument reset
Recall setup
Saved setup
1 (smuX.AUTORANGE_ON)
Usage
sourceAutorange
= smu
X
.source.autorange
Y
smu
X
.source.autorange
Y
=
sourceAutorange
sourceAutorange
The state of the source autorange control. Set to one of the following:
0
or
smu
X
.AUTORANGE_OFF
: Disables source autorange
1
or
smu
X
.AUTORANGE_ON
: Enables source autorange
X
Source-measure unit (SMU) channel (for example,
smua.source.autorangev
applies to SMU channel A)
Y
SMU source function (
v
= voltage,
i
= current)
Details
This attribute indicates the source autorange state. Its value is
smu
X
.AUTORANGE_OFF
when the
SMU source circuit is on a fixed range and
smu
X
.AUTORANGE_ON
when it is in autorange mode.
Setting this attribute to
smu
X
.AUTORANGE_OFF
puts the SMU on a fixed source range. The fixed
range used is the present SMU source circuit range.
Setting this attribute to
smu
X
.AUTORANGE_ON
puts the SMU source circuit into autorange mode. If
the source output is on, the SMU immediately changes range to the range most appropriate for the
value being sourced if that range is different than the present SMU range.
Autorange is disabled if the source level is edited from the virtual front panel. Setting the source
range also turns off autorange when set by using the
smu
X
.source.range
Y
attribute.
Resetting the instrument selects the
smu
X
.AUTORANGE_ON
.
Example
smua.source.autorangev = smua.AUTORANGE_ON
Enables volts source autorange for
SMU channel A.
Also see
(on page 7-224)