Section 7: TSP command reference
2606B System SourceMeter® Instrument Reference Manual
7-136
2606B-901-01 Rev. B / May 2018
lan.trigger[N].pseudostate
This attribute sets the simulated line state for the LAN trigger.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
LAN trigger N reset
Recall setup
Not saved
1
Usage
pseudostate
= lan.trigger[
N
].pseudostate
lan.trigger[
N
].pseudostate =
pseudostate
pseudostate
The simulated line state (
0
or
1
)
N
A number representing the LAN event number (1 to 8)
Details
This attribute can be set to initialize the pseudo line state to a known value.
Setting this attribute does not cause the LAN trigger to generate any events or output packets.
Example
print(lan.trigger[1].pseudostate)
Get the present simulated line state for the LAN
event 1.
Also see
None
lan.trigger[N].stimulus
This attribute specifies events that cause this trigger to assert.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
LAN trigger N reset
Recall setup
Not saved
0
Usage
triggerStimulus
= lan.trigger[
N
].stimulus
lan.trigger[
N
].stimulus =
triggerStimulus
triggerStimulus
The LAN event identifier used to trigger the event
N
A number specifying the trigger packet over the LAN for which to set or query
the trigger source (1 to 8)
Details
This attribute specifies which event causes a LAN trigger packet to be sent for this trigger. Set
triggerStimulus
to one of the trigger event IDs, which are shown in the following table.