Section 7: TSP command reference
2606B System SourceMeter® Instrument Reference Manual
7-74
2606B-901-01 Rev. B / May 2018
display.screen
This attribute contains the selected display screen.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
Recall setup
Saved setup
2 (display.SMUA_SMUB)
Usage
displayID
= display.screen
display.screen =
displayID
displayID
One of the following values:
0
or
display.SMUA
: Displays source-measure and compliance for SMU A
1
or
display.SMUB
: Displays source-measure and compliance for SMU B
2
or
display.SMUA_SMUB
: Displays source-measure for SMU A and SMU B
3
or
display.USER
: Displays the user screen
Details
Setting this attribute selects the display screen for the virtual front panel. This performs the same
action as selecting the DISPLAY key on the virtual front panel. The text for the display screen is set
by
display.settext()
.
Read this attribute to determine which of the available display screens was last selected.
Example
display.screen = display.SMUA
Selects the source-measure and compliance
limit display for SMU A.
Also see
(on page 7-76)
display.sendkey()
This function sends a code that simulates the action of a virtual front-panel control.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
display.sendkey(
keyCode
)
keyCode
A parameter that specifies the key press to simulate; see
Details
for more
information
Details
This command simulates pressing a virtual front-panel key or navigation wheel, or turning the
navigation wheel one click to the left or right.