20-20
Return to
2600AS-901-01 Rev. B / September 2008
Section 20: Calibration
Series 2600A System SourceMeter® Instruments Reference Manual
Figure 20-4
Connections for contact check 0
Ω
calibration
b.
Allow the readings to settle, then get the Series 2600A readings:
r0_hi, r0_lo = smua.contact.r()
c.
Connect
a
50
Ω
resistor between the sense low and output low terminals, as shown in
. Also connect a second 50
Ω
resistor between the sense high and output
high terminals as shown in the figure.
d.
Allow the readings to settle, then get the Series 2600A readings:
r50_hi, r50_lo = smua.contact.r()
e.
Send the contact check low calibration command:
smua.contact.calibratelo(r0_lo, Z_actual,r50_lo,50_ohm_actual)
Where:
r0_lo
-- Series 2600A 0W low measurement
Z_actual
-- actual zero value
r50_lo
-- Series 2600A 50W low measurement
50_ohm_actual -- actual 50W resistor value
Typical values:
smua.contact.calibratelo(r0_lo,0,r50_lo,50.15)
f.
(Where
r0_lo
and
r50_lo
are measurements taken in steps 2 and 4 above.)
g.
Send the contact check high calibration command:
smua.contact.calibratehi(r0_hi, Z_actual,r50_hi,50_ohm_actual)
Where:
r0_hi
-- Series 2600A 0W high measurement
Z_actual
-- actual zero value
r50_hi
-- Series 2600A 50W high measurement
50_ohm_actual -- actual 50W resistor value
Typical values:
smua.contact.calibratehi(r0_hi,0,r50_hi,50.15)
(Where
r0_hi
and
r50_hi
are measurements taken in steps 2 and 4 above.)
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
IEEE-488
LINE RATING
100-240VAC
50, 60Hz
240VA MAX.
DIGITAL I/O
RS-232
LINE FUSE
SLOWBLOW
3.15A, 250V
TSP-Link
R
CAT I
S
S
LO LO
HI
HI
G
G
G
G
CHANNEL A
!
LAN
NO AUTO-MDIX
!
!
CAT I
S
S
LO
LO
HI
HI
G G G
G
CHANNEL B
!
Model 2602A/2612A Channel A
Connections Shown
Output
LO
Sense
LO
Output
HI
Sense
HI