2600AS-901-01 Rev. B / September 2008
Return to
19-229
Series 2600A System SourceMeter® Instruments Reference Manual
Section 19: Remote Commands
tspnet.timeout
Attribute
Sets timeout value for
tspnet.connect(), tspnet.execute(),
and
tspnet.read()
commands.
Default
20
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage
tspnet.timeout = seconds_value
seconds_value = tspnet.timeout
seconds_value
Timeout value in seconds.
Remarks
• This setting sets the duration the
tspnet.connnect, tspnet.read,
and
tspnet.execute
commands will wait for a response. The time is specified in seconds. The
default value is 5.0 seconds. The timeout may contain fractional seconds but is only accurate to
the nearest 10mS. The timeout may be between 0.0 and 30 seconds.
• Errors:
Invalid Timeout
Example
tspnet.timeout = 10.0
tspnet.tsp.abort
Function
Aborts remote instrument execution.
TSP-Link
accessibility
This function can be accessed from a remote TSP-Link node.
Usage
tspnet.tsp.abort(connection_id)
connection_id
Integer value used as a handle for other
tspnet
commands.
Remarks
• This convenience command simply sends an “
abort
” string to the remote instrument.
• Errors:
Invalid Specified Connection
Connection Not Available
Write Failed
Example
tspnet.tsp.abort(myconnection)