19-210
Return to
2600AS-901-01 Rev. B / September 2008
Section 19: Remote Commands
Series 2600A System SourceMeter® Instruments Reference Manual
trigger.timer[N].stimulus
Replace N with the values 1-8
Attribute
Selects which event will start the timer.
Default
0
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage
eventid = trigger.timer[n].stimulus
-- Reads timer delay trigger event.
trigger.timer[n].stimulus = eventid
-- Writes timer delay trigger event.
n
Trigger timer number.
event_id
Event that triggers the timer delay.
Remarks
• This attribute selects which event starts the timer.
• Set this attribute to the event id of any trigger event to wait for that event.
• Use zero to disable event processing.
trigger.timer[N].wait
Replace N with the values 1-8
Attribute
Wait for a timer trigger.
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage
triggered = trigger.timer[n].wait(timeout)
n
Trigger timer number on which to wait.
timeout
Maximum time (in seconds) to wait for the trigger.
triggered
Trigger detection indicator.
Remarks
• This function will wait for a timer trigger.
• If the timer generated one or more trigger event since the last time
trigger.timer[n].wait
or
trigger.timer[n].clear
was called, this function will return immediately.
• After waiting for a trigger with this function, the event detector will be automatically reset and
rearmed. This is true regardless of the number of events detected.