Series 2600A System SourceMeter® Instruments Reference Manual
Section 19: Remote Commands
smuX.cal.lock
X = SMU channel (a or b)
Function
Disables commands that change calibration settings.
TSP-Link
accessibility
This function can be accessed from a remote TSP-Link node.
Usage
smuX.cal.lock()
Remarks
• This function will disable the calibration functions that can change the calibration settings. It is an
error to call this function while the calibration state is
smuX.CALSTATE_CALIBRATING
. The
calibration constants must be written to nonvolatile memory, or a previous calibration set must be
restored prior to locking calibration.
Details
See
.
Also see
Example
Disable calibration functions for SMU A:
smua.cal.lock()
smuX.cal.password
X = SMU channel (a or b)
Attribute
Password to enable calibration.
Default
“KI0026XX”
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage
smuX.cal.password = newpassword
newpassword
The new password (string).
Remarks
• A new password can only be assigned when calibration has been unlocked.
• The calibration password is write-only and cannot be read.
Details
See
.
Example
Assign a new calibration password for SMU A:
smua.cal.password = "LetMeIn"