2600AS-901-01 Rev. B / September 2008
Return to
19-81
Series 2600A System SourceMeter® Instruments Reference Manual
Section 19: Remote Commands
lan.trigger[N].connected
Replace N with values 1-8.
Attribute
The LAN event connection state.
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage
connected = lan.trigger[lanevent].connected
lanevent
The LAN event number.
connected
The LAN event connection state.
Remarks
• Set to
true
when the LAN trigger is connected and ready to send trigger events following a
successful
lan.trigger[n].connect
command. If the LAN trigger is not ready to send
trigger events, this value is set to
false
.
• Set to
false
when either
lan.trigger[n].protocol
or
lan.trigger[n].ipaddress
attributes are changed or the remote connection closes the connection.
lan.trigger[N].disconnect
Replace N with values 1-8.
Function
Disconnects the LAN trigger.
TSP-Link
accessibility
This function can be accessed from a remote TSP-Link node.
Usage
lan.trigger[lanevent].disconnect()
lanevent
The LAN event number.
Remarks
• For TCP connections, this closes the TCP connection.
• The LAN trigger will automatically disconnect when either the
lan.trigger[n].protocol
or
lan.trigger[n].ipaddress
attributes for this event are changed.
lan.trigger[N].EVENT_ID
Replace N with values 1-8.
Attribute
Indicates the trigger event detector LAN event number.
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage
event_id = lan.trigger[lanevent].EVENT_ID
event_id
The trigger event number.
lanevent
The LAN event number.
Remarks
• Set the stimulus of any trigger event detector to the value of this constant to have it respond to