Series 2600A System SourceMeter® Instruments Reference Manual
Section 19: Remote Commands
digio.trigger[N].release
Replace N with the number of the digital I/O trigger line: 1 to 14.
Function
Releases an indefinite length or latched trigger
.
TSP-Link
accessibility
This function can be accessed from a remote TSP-Link node.
Usage
digio.trigger[n].release()
n
The trigger line.
Remarks
• Releases a trigger that was asserted with an indefinite pulse width, as well as a trigger that was
latched in response to receiving a synchronous mode trigger.
Details
See
.
Also see
Example
Releases trigger line 4
:
digio.trigger[4].release()
digio.trigger[N].stimulus
Replace N with the number of the digital I/O trigger line: 1 to 14.
Attribute
Selects the event used to generate a trigger.
Default
0
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage
stimulus = digio.trigger[n].stimulus
-- Reads stimulus event.
digio.trigger[n].stimulus = stimulus
-- Writes stimulus event.
n
The number of the trigger line.
stimulus
The identifier for the triggering event.
Remarks
• Use this attribute to select an event that triggers the digital output line.
• Set this attribute to 0 (zero) to disable automatic trigger output.
• Do not use the
stimulus
attribute for generating output triggers under script control. Use
digio.trigger[n].assert
instead.