F-10
Measurement Considerations
Model 2520 User’s Manual
Magnetic coupling
Magnetic coupling between sense and current leads can also affect measurements.
Figure F-9A shows connections that will result in considerable coupling due to the rela-
tively long exposed signal lead area. Carrying the sense shields as close to the DUT as
possible (Figure F-9B) will reduce magnetic coupling and improve response. Twisting the
exposed sense leads and even the cables themselves may be necessary to reduce coupling
to an acceptable level.
Figure F-9
Magnetic coupling
DUT
Connecting
Cables
Coupling
Caused by
Exposed Leads
A. Coupling caused by exposed leads.
DUT
Connecting
Cables
Coupling
Reduced by
Shielding
B. Coupling reduced by shielding.
Voltage Sense HI
Current Output HI
Current Output LO
Voltage Sense LO
Voltage Sense HI
Current Output HI
Current Output LO
Voltage Sense LO
Summary of Contents for 2520
Page 250: ...A Specifications ...
Page 254: ...B StatusandErrorMessages ...
Page 262: ...C DataFlow ...
Page 266: ...D IEEE 488BusOverview ...
Page 281: ...E IEEE 488andSCPI ConformanceInformation ...
Page 285: ...F MeasurementConsiderations ...
Page 308: ...G GPIB488 1Protocol ...
Page 312: ...H ExamplePrograms ...
Page 317: ...I ContinuousPulseMode ...