Model 2520 User’s Manual
SCPI Command Reference
14-57
Configure sample buffer
POINts <n>
:TRACe:POINts <n>
Specify number of samples
Parameters
<n> =
1 to 3000
Specify number of samples
MINimum
1
MAXimum
32000
DEFault
100
Query
:POINts?
Query number of samples
:POINts? MINimum
Query smallest allowable sample number
:POINts? MAXimum
Query largest allowable sample number
:POINts? DEFault
Query *RST default sample number
Description
This command sets the number of samples to return for the
TRAC:DATA? query. Every time the pulse width
(SOUR:PULS:WIDTh) is changed, the TRAC:POINts value defaults to
either the (Pulse Width
×
10e6) in DC mode, or (Pulse Width
×
10e6 +
5) in pulse mode, or 3000, whichever is less. For example, if the pulse
width is 10
µ
s, the number of points will default to 100 in DC mode, or
105 in the pulse mode.
Trigger subsystem
The Trigger subsystem is made up of a series of commands and subsystems to configure
the Trigger Model. These commands and subsystems are summarized in
NOTE
See
for more details on triggering and the trigger model.
Initiate source/measure cycle
INITiate
:INITiate[:IMMediate]
Take Model 2520 out of idle state
Description
This command is used to initiate source and measure operation by tak-
ing the Model 2520 out of idle. The :READ? and :MEASure? com-
mands also perform an initiation.
Command processing will halt until the trigger model returns to the idle
state, with the following exceptions:
•
ABORt
•
SDC
•
DCL
•
IFC
•
SYSTem:PRESet, *RST, or *RCL
•
*TRG or GET with TRIG:SOUR BUS
If any other command is sent after an INIT, only SDC, IFC, and DCL
will work.
Summary of Contents for 2520
Page 250: ...A Specifications ...
Page 254: ...B StatusandErrorMessages ...
Page 262: ...C DataFlow ...
Page 266: ...D IEEE 488BusOverview ...
Page 281: ...E IEEE 488andSCPI ConformanceInformation ...
Page 285: ...F MeasurementConsiderations ...
Page 308: ...G GPIB488 1Protocol ...
Page 312: ...H ExamplePrograms ...
Page 317: ...I ContinuousPulseMode ...