14-34
SCPI Command Reference
Model 2520 User’s Manual
For example, if you expect to source levels around 300mA, send the fol-
lowing command:
:SOURce1:CURRent:RANGe 0.3
The above command will select the 500mA range for the laser diode
current source.
As listed in
Parameters
, you can also use the MINimum, MAXimum
and DEFault parameters to manually select the source range. The UP
parameter selects the next higher source range, while DOWN selects the
next lower source range.
Set amplitudes
[:IMMediate][:AMPLitude] <n>
:SOURce[1]:CURRent[:LEVel][:IMMediate][:AMPLitude] <n> Set fixed current source amplitude
Parameters
<n> =
0 to 5.0
Set source amplitude (amps)
DEFault
0A
MINimum
0A
MAXimum
5.0A
Query
:CURRent?
Query programmed source amplitude
:CURRent? DEFault
Query *RST default amplitude
:CURRent? MINimum
Query lowest allowable amplitude
:CURRent? MAXimum
Query highest allowable amplitude
Description
This command is used to immediately update the amplitude of a fixed
source for both DC and pulse functions (with the pulse function, the
amplitude is the high pulse level). This command is not valid if the list
or sweep mode is selected.
NOTE
The sourcing :MODE command is used to select a fixed source. See
The specified amplitude cannot exceed the selected source range. For
example, if the source is on the 500mA range, you will not be able to set
the source amplitude to 1A.
NOTE
For the DC function, the current cannot be set above 1A.
LOW <n>
:SOURce[1]:CURRent[:LEVel]:LOW <n>
Set current pulse low amplitude
Parameters
<n> =
0 to 0.150
Set pulse low amplitude (amps)
DEFault
0A
MINimum
0A
MAXimum
0.150A
Summary of Contents for 2520
Page 250: ...A Specifications ...
Page 254: ...B StatusandErrorMessages ...
Page 262: ...C DataFlow ...
Page 266: ...D IEEE 488BusOverview ...
Page 281: ...E IEEE 488andSCPI ConformanceInformation ...
Page 285: ...F MeasurementConsiderations ...
Page 308: ...G GPIB488 1Protocol ...
Page 312: ...H ExamplePrograms ...
Page 317: ...I ContinuousPulseMode ...