Model 2520 User’s Manual
SCPI Command Reference
14-31
[:STATe] <b>
[:SENSe[1]]:AVERage:[:STATe] <b>
Enable/disable digital filter
:SENSe2:AVERage:[:STATe] <b>
Enable/disable digital filter
:SENSe3:AVERage:[:STATe] <b>
Enable/disable digital filter
Parameters
<b> =
0 or OFF
Disable digital filter
1 or ON
Enable digital filter
Query
[:STATe]?
Query state of average digital filter
Description
These commands are used to enable or disable the average digital filter.
When enabled voltage and current readings are filtered according to
how may readings are averaged (see
disabled, the digital filter stage is bypassed.
Abort sweep
:CABort[:LEVel] <n>
:SENSe2:SWEep:CABort[:LEVel] <n>
Set detector #1 sweep abort current
:SENSe3:SWEep:CABort[:LEVel] <n>
Set detector #2 sweep abort current
Parameters
<n> =
0 to 0.105
Sweep abort current in amps
DEFault
0.105
MINimum
0
MAXimum
0.105
Query
:CABort[:LEVel]?
Query sweep abort current threshold
Description
These commands provide means to set a maximum detector current for
detector #1 or detector #2 that will cause a sweep to be aborted if the
corresponding detector current threshold is reached. Sweep abort must
be enabled with the :STATe command (see below).
The Sweep Aborted bit (Bit 7) in the Measurement Event Register will
be set if the sweep aborts early because of reaching the maximum detec-
tor current threshold for a particular channel. The Model 2520 can be
programmed to generate an SRQ under these conditions. See Section 11
for register and SRQ details.
:CABort:STATe <b>
:SENSe2:SWEep:CABort:STATe <b>
Enable/disable detector #1 sweep abort
:SENSe3:SWEep:CABort:STATe <b>
Enable/disable detector #2 sweep abort
Parameters
<b> =
0 or OFF
Disable sweep abort
1 or ON
Enable sweep abort
Query
:STATe?
Query sweep abort state
Description
These commands enable or disable sweep abort when the programmed
current threshold is reached (see [:LEVel] above).
Summary of Contents for 2520
Page 250: ...A Specifications ...
Page 254: ...B StatusandErrorMessages ...
Page 262: ...C DataFlow ...
Page 266: ...D IEEE 488BusOverview ...
Page 281: ...E IEEE 488andSCPI ConformanceInformation ...
Page 285: ...F MeasurementConsiderations ...
Page 308: ...G GPIB488 1Protocol ...
Page 312: ...H ExamplePrograms ...
Page 317: ...I ContinuousPulseMode ...