Model 2520 User’s Manual
SCPI Command Reference
14-17
DATA?
:CALCulate[1]:DATA?
Read laser diode math (CALC1) result
:CALCulate2:DATA?
Read detector #1 math (CALC2) result
:CALCulate3:DATA?
Read detector #2 math (CALC3) result
:CALCulate4:DATA?
Read delta math (CALC4) result
Description
These query commands are used to read the result of the CALC1,
CALC2, CALC3, or CALC4 calculation. The largest valid calculation
result can be ±9.9e37, which (defined by SCPI) is infinity.
An invalid NAN (not a number) result of +9.91e37 indicates that
CALC1, CALC2, CALC3, or CALC4 is disabled or that a division by
zero error has occurred. See
LATest?
:CALCulate[1]:DATA:LATest?
Read latest CALC1 result
:CALCulate2:DATA:LATest?
Read latest CALC2 result
:CALCulate3:DATA:LATest?
Read latest CALC3 result
:CALCulate4:DATA:LATest?
Read latest CALC4 result
Description
These commands work exactly like CALC1:DATA?, CALC2:DATA?,
CALC3:DATA?, and CALC4:DATA? except that they return only the
latest CALC1, CALC2, CALC3, or CALC4 result.
DISPlay subsystem
The display subsystem controls the display of the Model 2520 and is summarized in
Control display
ENABle <b>
:DISPlay:ENABle <b>
Control display circuitry
Parameters
<b> =
0 or OFF
Disable display circuitry
1 or ON
Enable display circuitry
Query
:ENABle?
Query state of display
Description
This command is used to enable and disable the front panel display cir-
cuitry. When disabled, the instrument operates at a higher speed. While
disabled, the display is frozen with the following message:
FRONT PANEL DISABLED
Press LOCAL to resume.
As reported by the message, all front panel controls (except LOCAL
and OUTPUT OFF) are disabled. Normal display operation can be
resumed by using the :ENABle command to enable the display or by
putting the Model 2520 into local.
Summary of Contents for 2520
Page 250: ...A Specifications ...
Page 254: ...B StatusandErrorMessages ...
Page 262: ...C DataFlow ...
Page 266: ...D IEEE 488BusOverview ...
Page 281: ...E IEEE 488andSCPI ConformanceInformation ...
Page 285: ...F MeasurementConsiderations ...
Page 308: ...G GPIB488 1Protocol ...
Page 312: ...H ExamplePrograms ...
Page 317: ...I ContinuousPulseMode ...