8-10
Triggering
Model 2520 User’s Manual
Figure 8-6
Trigger model (remote operation)
Turn On Outputs
Idle
PULSE
DELAY
Start
Sweep
No
= Output Trigger
✛
Yes
No
✛
1
✛
= GPIB Default
TRIGger:OUTPut
Yes
Sweep
Complete
?
Another
Trigger
?
Filter
Process
(Repeat)
Trigger Event
Detector
Trig-In
Event
Trigger
Layer
Note: The following commands
place the Model 2520 into
idle: DCL, SDC, ABORt,
*RST, SYSTem:PREset, and
*RCL.
TRIGger:SOURce
✛
IMMediate
BUS
TIMer
MANual
TLINK
NSTest
PSTest
BSTest
No
See Note
INITiate
?
NONE
|
TRIGger
TRIGger:COUNt
<n>
|
INF
SOURce1:PULSe:WIDTh
✛
10
µ
sec
SOURce1:PULSe:DELay
✛
1
µ
sec
Note: Sweep performed only if
enabled. Filter count
determines filter process
repeat cycles.
Summary of Contents for 2520
Page 250: ...A Specifications ...
Page 254: ...B StatusandErrorMessages ...
Page 262: ...C DataFlow ...
Page 266: ...D IEEE 488BusOverview ...
Page 281: ...E IEEE 488andSCPI ConformanceInformation ...
Page 285: ...F MeasurementConsiderations ...
Page 308: ...G GPIB488 1Protocol ...
Page 312: ...H ExamplePrograms ...
Page 317: ...I ContinuousPulseMode ...