Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 8: TSP command reference
2461-901-01 A/November 2015
8-295
Example
digio.line[3].mode = digio.MODE_TRIGGER_OUT
trigger.model.setblock(5, trigger.BLOCK_NOTIFY, trigger.EVENT_NOTIFY2)
trigger.digout[3].stimulus = trigger.EVENT_NOTIFY2
Define trigger model block 5 to be the notify 2 event. Assign the notify 2 event to be the stimulus for digital
output line 3.
Also see
(on page 3-112)
trigger.model.setblock() —
trigger.BLOCK_RESET_BRANCH_COUNT
This function creates a block in the trigger model that resets a branch counter to 0.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Restore configuration
Instrument reset
Power cycle
Configuration script
Not applicable
Usage
trigger.model.setblock(
blockNumber
, trigger.BLOCK_RESET_BRANCH_COUNT,
counter
)
blockNumber
The sequence of the block in the trigger model
counter
The block number of the counter that is to be reset
Details
When the trigger model reaches the Counter Reset block, it resets the count of the specified Branch
on Counter block to zero.