Section 8: TSP command reference
Model 2461 Interactive SourceMeter® Instrument Reference Manual
8-284
2461-901-01 A/November 2015
Example 2
trigger.model.load("Empty")
trigger.model.setblock(1, trigger.BLOCK_CONFIG_RECALL, "measTrigList")
trigger.model.setblock(2, trigger.BLOCK_BUFFER_CLEAR)
trigger.model.setblock(3, trigger.BLOCK_CONFIG_NEXT, "measTrigList")
print(trigger.model.getblocklist())
Clear the trigger model.
Recall index 1 of a configuration list named
measTrigList
.
Clear reading buffer named
defbuffer1
.
Recall the second index of a configuration list named
measTrigList
.
Output:
1) CONFIG_RECALL CONFIG_LIST: measTrigList INDEX: 1
2) BUFFER_CLEAR BUFFER: defbuffer1
3) CONFIG_NEXT CONFIG_LIST: measTrigList
Example 3
trigger.model.setblock(7, trigger.BLOCK_CONFIG_NEXT, "measTrigList",
"sourTrigList")
Configure trigger block 7 to load the next index in both the configuration list named
measTrigList
and the
configuration list named
sourTrigList
.
Also see
(on page 3-30)
trigger.model.setblock() — trigger.BLOCK_CONFIG_PREV
This function defines a trigger model block that recalls the settings stored at the previous index in a source or
measure configuration list, or both a source and measure configuration list.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Restore configuration
Instrument reset
Power cycle
Configuration script
Not applicable
Usage
trigger.model.setblock(
blockNumber
, trigger.BLOCK_CONFIG_PREV,
configurationList
)
trigger.model.setblock(
blockNumber
, trigger.BLOCK_CONFIG_PREV,
configurationList
,
optionalConfigList)
blockNumber
The sequence of the block in the trigger model
configurationList
A string that defines the source or measure configuration list to
recall
optionalConfigList
A string that defines the second configuration list to recall the
index from; must be the opposite type of list than the first; for
example, if the first configuration list is a measure list, the
second configuration list must be a source list