Section 8: TSP command reference
Model 2461 Interactive SourceMeter® Instrument Reference Manual
8-270
2461-901-01 A/November 2015
trigger.model.load() — SimpleLoop
This function loads a predefined trigger model configuration that makes a specific number of measurements.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
trigger.model.load("SimpleLoop",
count
)
trigger.model.load("SimpleLoop",
count
,
sDelay
)
trigger.model.load("SimpleLoop",
count
,
sDelay
,
bufferName
)
trigger.model.load("SimpleLoop",
count
,
sDelay
,
bufferName, readingBlock
)
count
The number of measurements the instrument will make
sDelay
The delay time before each measurement (167 ns to 10 ks); default is 0 for
no delay
bufferName
Indicates the reading buffer to use; the default buffers (
defbuffer1
or
defbuffer2
) or the name of a user-defined buffer; if no buffer is
specified,
defbuffer1
is used
readingBlock
Define a measure or digitize block for the trigger model; options are:
•
trigger.READING_ACTIVE
: Add a measure or digitize block to the
trigger model based on the active function; if no option defined,
trigger.READING_ACTIVE
is used
•
trigger.READING_MEASURE
: Adds a measure block to the trigger
model
•
trigger.READING_DIGITIZE
: Adds a digitize block to the trigger
model
Details
This command sets up a loop that sets a delay, makes a measurement, and then repeats the loop the
number of times you define in the count parameter.
After selecting a trigger model template, you can view the trigger model blocks in a graphical format
by pressing the front-panel
MENU
key and under Trigger, selecting
Configure
. You can also add or
delete blocks and change trigger model settings from this screen. You can use the
trigger.model.getblocklist()
command to view the trigger model blocks in a list format.