Section 8: TSP command reference
Model 2461 Interactive SourceMeter® Instrument Reference Manual
8-186
2461-901-01 A/November 2015
Example
smu.source.func = smu.FUNC_DC_VOLTAGE
smu.source.level = 1
Set the instrument to source
voltage and set it to source 1 V.
Also see
(on page 8-184)
(on page 8-188)
(on page 8-188)
smu.source.offmode
This attribute defines the state of the source when the output is turned off.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW) Yes
Restore configuration
Instrument reset
Power cycle
Source configuration list
Function change
Configuration script
Source configuration list
smu.OFFMODE_NORMAL
Usage
sourceOffMode
= smu.source.offmode
smu.source.offmode =
sourceOffMode
sourceOffMode
The output-off setting; set to one of the following values (see the
Details
below for
specifics regarding each option):
•
smu.OFFMODE_NORMAL
•
smu.OFFMODE_ZERO
•
smu.OFFMODE_HIGHZ
•
smu.OFFMODE_GUARD
Details
When the Model 2461 is set to the normal output-off state, the following settings are made when the
source is turned off:
•
The measurement sense is set to 2-wire
•
The voltage source is selected and set to 0 V
•
The current limit is set to 10 % of the full scale of the present measurement function autorange
value
•
If source readback is off, Output Off is displayed in the Home screen Source area
•
If source readback is on, the actual measurement is displayed in the Home screen Source area
•
If measurement is set to resistance, dashes (--.----) are shown in the Home screen Source area
•
The Source button on the Home screen shows the value that will be sourced when the output is
turned on again
When the high-impedance output-off state is selected and the output is turned off:
•
The measurement sense is set to 2-wire
•
The output relay opens, disconnecting the instrument as a load