Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 8: TSP command reference
2461-901-01 A/November 2015
8-107
smu.digitize.count
This attribute sets the number of measurements to digitize when a measurement is requested.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Configuration script
10,000
Usage
count
= smu.digitize.count
smu.digitize.count =
count
count
The number of measurements to make (1 to 55,000,000)
Details
The digitizer makes the number of readings set by this command in the time set by the sample rate.
This command does not affect the trigger model.
Example
smu.digitize.func = smu.FUNC_DIGITIZE_CURRENT
smu.digitize.aperture = smu.APERTURE_AUTO
smu.digitize.samplerate = 1000000
smu.digitize.count = 10
print(smu.digitize.read())
Set the digitize function to measure current.
Set the sample rate to 1,000,000, with a
count of 10, and automatic aperture.
Make a digitize measurement.
Example output:
-0.0039799990218
Also see
(on page 2-113)
(on page 8-106)
(on page 8-128)
smu.digitize.displaydigits
This attribute describes the number of digits that are displayed on the front panel for the selected function.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Configuration script
Measure configuration list
smu.DIGITS_4_5
Usage
value
= smu.digitize.displaydigits
smu.digitize.displaydigits =
value
value
•
6½ display digits:
smu.DIGITS_6_5
•
5½ display digits:
smu.DIGITS_5_5
•
4½ display digits:
smu.DIGITS_4_5
•
3½ display digits:
smu.DIGITS_3_5