Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 8: TSP command reference
2461-901-01 A/November 2015
8-101
Details
The
scriptVar
.save()
function saves a script to nonvolatile memory or a USB flash drive. The
root folder of the USB flash drive has the absolute path
/usb1/
.
If no
filename
is specified, the script is saved to internal nonvolatile memory. If a
filename
is
given, the script is saved to the USB flash drive.
If you set
scriptVar
to
autoexec
, the script is run when the instrument powers up. You must
delete the existing
autoexec
script before saving the new one. Note that performing a system reset
does not delete the
autoexec
script.
You can add the file extension, but it is not required. The only allowed extension is
.tsp
(see
Example 2).
Example 1
test8.save()
Saves the script referenced by the variable
test8
to nonvolatile memory.
Example 2
test8.save("/usb1/myScript.tsp")
Saves the script referenced by the variable
test8
to a file named
myScript.tsp
on your
USB flash drive.
Also see
(on page 7-5)
scriptVar.source
This attribute contains the source code of a script.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
No
Not applicable
Not applicable
Not applicable
Usage
code
=
scriptVar
.source
code
The body of the script
scriptVar
The name of the variable that references the script that contains the source code
Details