Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 8: TSP command reference
2461-901-01 A/November 2015
8-97
reset()
This function resets commands to their default settings and clears the buffers.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
No
Usage
reset()
reset(
system
)
system
If the node is the master, the entire system is reset:
true
Only the local group is reset:
false
Details
The
reset()
command in its simplest form resets the entire TSP-enabled system, including the
controlling node and all subordinate nodes.
If you want to reset a specific instrument, use the
node[
N
].reset()
command. Also use the
node[
N
].reset()
command to reset an instrument on a subordinate node.
When no value is specified for
system
, the default value is
true
.
You can only reset the entire system using
reset(true)
if the node is the master. If the node is not
the master node, executing this command generates an error event.
Example
reset(true)
If the node is the master node, the entire system is
reset; if the node is not the master node, an error
event is generated.
Also see
(on page 2-141)
localnode.prompts4882
This attribute enables and disables the generation of prompts for IEEE Std 488.2 common commands.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Power cycle
Not saved
localnode.ENABLE
Usage
prompting
= localnode.prompts4882
localnode.prompts4882 =
prompting
prompting
IEEE Std 488.2 prompting mode:
•
Disable prompting:
localnode.DISABLE
•
Enable prompting:
localnode.ENABLE