Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 8: TSP command reference
2461-901-01 A/November 2015
8-55
Example
digio.writeport(63)
Sets digital I/O lines 1 through 6 high (binary
111111).
Also see
(on page 8-53)
Digital I/O port configuration
(on page 3-87)
display.changescreen()
This function changes which front-panel screen is displayed.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
display.changescreen(s
creenName
)
screenName
The screen to display:
•
Home screen:
display.SCREEN_HOME
•
Home screen with large readings:
display.SCREEN_HOME_LARGE_READING
•
Reading table screen:
display.SCREEN_READING_TABLE
•
Graph screen (opens last selected tab):
display.SCREEN_GRAPH
•
Histogram:
display.SCREEN_HISTOGRAM
•
GRAPH swipe screen:
display.SCREEN_GRAPH_SWIPE
•
SETTINGS swipe screen:
display.SCREEN_SETTINGS_SWIPE
•
SOURCE swipe screen:
display.SCREEN_SOURCE_SWIPE
•
STATISTICS swipe screen:
display.SCREEN_STATS_SWIPE
•
USER swipe screen:
display.SCREEN_USER_SWIPE
Example
display.clear()
display.changescreen(display.SCREEN_USER_SWIPE)
display.settext(display.TEXT1, "Batch A122")
display.settext(display.TEXT2, "Test running")
Clear the USER swipe screen and switch to display the USER swipe screen.
Set the first line to read "Batch A122" and the second line to display "Test running":
Also see
(on page 8-66)