Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 6: SCPI command reference
2461-901-01 A/November 2015
6-191
Example
:TRIG:BLOC:BRAN:EVEN 6, DISP, 2
When the trigger model reaches this block, if the front-panel TRIGGER key has been pressed, the trigger
model returns to block 2. If the TRIGGER key has not been pressed, the trigger model continues to block 7
(the next block in the trigger model).
Also see
(on page 3-118)
:TRIGger:BLOCk:BRANch:LIMit:CONStant
This command defines a trigger model block that goes to a specified block if a measurement meets preset criteria.
Type
Affected by
Where saved
Default value
Command only
Recall settings
Instrument reset
Power cycle
Save settings
Not applicable
Usage
:TRIGger:BLOCk:BRANch:LIMit:CONStant <blockNumber>, <limitType>, <limitA>,
<limitB>, <branchToBlock>
:TRIGger:BLOCk:BRANch:LIMit:CONStant <blockNumber>, <limitType>, <limitA>,
<limitB>, <branchToBlock>, <measureBlock>
<blockNumber>
The sequence of the block in the trigger model
<limitType>
The type of limit (
ABOVe
,
BELow
,
INside
, or
OUTside
)
<limitA>
The limit that the measurement is tested against; if
limitType
is set to:
•
ABOVe
: This value is ignored
•
BELow
: The measurement must be below this value
•
INside
: The low limit that the measurement is compared against
•
OUTside
: The low limit that the measurement is compared against
<limitB>
The upper limit that the measurement is tested against; if
limitType
is set to:
•
ABOVe
: The measurement must be above this value
•
BELow
: This value is ignored
•
INside
: The high limit that the measurement is compared against
•
OUTside
: The high limit that the measurement is compared against
<branchToBlock>
The block number of the trigger model block to execute when the measurement
meets the defined criteria
<measureBlock>
The block number of the measure or digitize block that makes the measurements to
be compared; if this is
0
or undefined, the trigger model uses the previous measure
or digitize block
Details
The branch-on-constant-limits block goes to a branching block if a measurement meets the criteria
set by this command.
The type of limit can be:
•
Above: The measurement is above the value set by limit B; limit A must be set, but is ignored
when this type is selected
•
Below: The measurement is below the value set by limit A; limit B must be set, but is ignored
when this type is selected