Section 6: SCPI command reference
Model 2461 Interactive SourceMeter® Instrument Reference Manual
6-32
2461-901-01 A/November 2015
:CALCulate2:<function>:LIMit<Y>:STATe
This command enables or disables a limit test on the measurement from the selected measure function.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
0 (OFF)
Usage
:CALCulate2:<function>:LIMit<Y>:STATe <state>
:CALCulate2:<function>:LIMit<Y>:STATe?
<function>
The measure function:
•
Current:
CURRent[:DC]
•
Resistance:
RESistance
•
Voltage:
VOLTage[:DC]
The digitize function:
•
Digitize current:
DIGitize:CURRent
•
Digitize voltage:
DIGitize:VOLTage
<Y>
Limit number:
1
or
2
<state>
Disable the limit test:
OFF
or
0
Enable the limit test:
ON
or
1
Details
This command enables or disables a limit test for the selected measurement function. When this
attribute is enabled, the limit
Y
testing occurs on each measurement made by the instrument. Limit
Y
testing compares the measurements to the high and low limit values. If a measurement falls outside
these limits, the test fails.
If you send this command without the
<function>
parameter, it sets the state of math operations
for all math functions. It will not change the setting for a digitize function.