Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 3: Functions and features
2461-901-01 A/November 2015
3-129
Starting the trigger model
Using the front panel:
1. Press the front-panel
TRIGGER
key for 2 s. A dialog box displays the available trigger methods.
The presently selected method is highlighted.
2. Select
Initiate Trigger Model
.
3. If the instrument is controlled remotely, a confirmation screen is displayed. Select
Yes
to change
to front-panel control and start the trigger model.
Using SCPI commands:
Send the command:
:INITiate
Using TSP commands:
Send the command:
trigger.model.initiate()
Aborting the trigger model
You can stop the trigger model while it is in progress. When you stop the trigger model, all trigger
model commands on the instrument are terminated, including sweeps.
Using the front panel:
Press the
TRIGGER
key for 2 s and select
Abort Trigger Model
.
Using SCPI commands:
Send the command:
:ABORt
Using TSP commands:
Send the command:
trigger.model.abort()